Electrodes for Laboratory
| LTG0,1/22, LTG0,1/23 | LTG1/22, LTG1/23 |
Conductivity cells
| LTC0,35/23 | LTC1/22, LTC1/23 || LTG0,1/22, LTG0,1/23 | LTG1/22, LTG1/23 |
LTC0,35/23
|
LTC1/22 or LTC1/23
|
|
Type | LTC0,35/23 | LTC1/22 or LTC1/23 |
Shaft | plastic, Ø 12 mm | plastic, Ø 12 mm |
Shaft length | 120 mm | 120 mm |
Electrode material | graphite | graphite |
Temperature sensor | Pt 1000 | LTC1/22: Pt100 LTC1/23: Pt1000 |
Measuring principle | 4-electrode technology | 2-electrode technology |
Cell constant | 0.33 cm-1 ± 5 % | 1 cm-1 ± 20 % |
Measuring range | 1 µS/cm ... 500 mS/cm | 1 µS/cm ... 100 mS/cm |
Temperature range | -5 ... 80 °C | -5 ... 80 °C |
Minimum immersion depth |
20 mm | 15 mm |
Application | drinking water, surface water, waste water, sea water industrial process water, acids, bases | drinking water, surface water, waste water, sea water |
Electrical connection | fixed cable | fixed cable |
Order key | LTC0,35/23 ... | LTC1/22 ..., LTC1/23 ... |
LTG0,1/22 or LTG0,1/23
|
LTG1/22 or LTG1/23
|
|
Type | LTG0,1/22 or LTG0,1/23 | LTG1/22 or LTG1/23 |
Shaft | glass, Ø 12 mm | glass, Ø 12 mm |
Shaft length | 120 mm | 120 mm |
Electrode material | platinum (8x17 mm) | platinum (3x4 mm) |
Temperature sensor | LTG0,1/22: Pt100 LTG0,1/23: Pt1000 |
LTG1/22: Pt100 LTG1/23: Pt1000 |
Measuring principle | 2-electrode technology | 2-electrode technology |
Cell constant | 0.1 cm-1 ± 20 % | 1 cm-1 ± 20 % |
Measuring range | 0.1 ... 200 µS/cm | 1 µS/cm ... 100 mS/cm |
Temperature range | -5 ... 100 °C | -5 ... 100 °C |
Minimum immersion depth |
35 mm | 25 mm |
Application | low ionic mediums (distillation plants, ion exchanger) | drinking water, surface water, waste water, sea water, swimming pools, beverages |
Electrical connection | fixed cable | fixed cable |
Order key | LTG0,1/22 ..., LTG0,1/23 ... | LTG1/22 ..., LTG1/23 ... |